The VicoMeter WKT-A1 Entry-Level Karl Fischer trace moisture analyzer uses the coulometric Karl Fischer (coulometry) principle to rapidly measure trace moisture, detecting water down to 1 ppm and determining trace moisture in solids, liquids, and powders within one minute. It offers a friendly interface with a 128×64 blue LCD and full Chinese menus. The detection circuit uses digital processing to overcome the temperature drift and power-supply fluctuation that troubled traditional instruments, improving testing precision and sensitivity. The whole instrument is built with a relay-free process design, eliminating the hidden risk of relay contact oxidation and extending service life. It is widely used in chemicals, petroleum, pharmaceuticals, power, food, research, and testing.
Technical Specifications
The VicoMeter WKT-A1 Entry-Level Karl Fischer trace moisture analyzer uses the coulometric Karl Fischer (coulometry) principle to rapidly measure trace moisture, detecting water down to 1 ppm and determining trace moisture in solids, liquids, and powders within one minute. It offers a friendly interface with a 128×64 blue LCD and full Chinese menus. The detection circuit uses digital processing to overcome the temperature drift and power-supply fluctuation that troubled traditional instruments, improving testing precision and sensitivity. The whole instrument is built with a relay-free process design, eliminating the hidden risk of relay contact oxidation and extending service life. It is widely used in chemicals, petroleum, pharmaceuticals, power, food, research, and testing.
| Model | WKT-A1 |
| Measuring Principle | Karl Fischercoulometric(coulometric) |
| Measurement Speed | 33.3 μg H2O/s (max) |
| Measuring Range | 2 μg H2O to 120 mg H2O |
| Measurement Accuracy | Error ≤ ±2 μg for water content below 10 μg; error ≤ ±5 μg for 10–500 μg water (injection error excluded); standard deviation ≤ 0.2% |
| Resolution | 0.1 μg H2O |
| Electrolysis Current | 0-400mA |
| Water Content (Concentration Range) | 1ppm-1000000ppm(0.0001%-100%) |
| Stirring Speed | Stepless knob adjustment |
| Drift/Blank Subtraction | Automatic subtraction |
| Special Functions | Drift coefficient correction, electrolysis rate adjustment, and self-diagnostic fault alerts |
| Display & Operation | 128×64 blue LCD |
| Calendar/Clock | Current time, date, and test time/date display |
| Endpoint Indication | Screen prompt / audible alarm / end-point lamp |
| Operating Environment | Temperature 2 ℃–50 ℃, humidity <90% |
| Power Consumption | ≤30W |
| Unit Dimensions | 300mm*205mm*110mm |
| Power Supply | 220V 50HZ 150W |
| Weight | 5kg |
Applicable Industries
| Liquid | Chemical industry: alcohols, ethers, acids, benzenes, phenols, organic solvents, and other products suited to Karl Fischer determination. Petroleum and power: insulating oil, transformer oil, and other oils. Pharmaceutical industry: pharmaceutical raw materials, etc. Pesticide industry: emulsifiers, etc. Other industries: lithium battery electrolytes, etc. |
| Solid | Freely soluble solids such as various inorganic salts and citric acid |
| Gas | Gases such as nitrogen, SF6, natural gas, liquefied petroleum gas, Freon, and butadiene |
| The following industries require a Karl Fischer oven | |
| Insoluble Solids | Lithium batteries, plastics, medical, ceramic powders, metal powders, etc. |
| Other Insoluble Solids | Insoluble solids, easily contaminated and cross-contaminating samples (lubricating greases, aldehydes and ketones), samples that react with Karl Fischer reagent, and samples with slow moisture release |