The electrode is the key component; surface contamination cuts sensitivity. Long-term use on oily samples makes electrodes sluggish at the endpoint, consuming excess reagent—cleaning is then required. Contamination is invisible but sluggish responses directly affect accuracy; electrodes must be cleaned periodically, though often overlooked.
For heavy contamination, wipe with acetone-dampened paper (reuse only after full evaporation), soak in dilute nitric acid for 24 hours then rinse and dry, or reactivate with potassium dichromate for 1 minute. A quick fix exists when time is short.
Rub both ends gently with very fine sandpaper and wipe clean for immediate improvement. When idle, drain all reagent from tubes, lines, and the reaction vessel to prevent crystallization blockage. Operators who simply power off after testing show insufficient awareness that maintenance and accuracy go hand in hand.
VicoMeter is an expert in moisture determination, rubber & plastics testing and petrochemical testing, offering a range of reliable, easy-to-use analytical and inspection instruments including rapid moisture analyzers, halogen moisture analyzers, Karl Fischer moisture analyzers, online moisture analyzers, DSC differential scanning calorimeters, HDT/Vicat softening point testers, melt flow indexers and impact testers. Whether for professional laboratory analysis, production workshops or rapid on-site testing, VicoMeter provides professional, targeted solutions meeting the quality control, production and in-line control needs of the chemical, pharmaceutical, plastics, food, lithium battery, petroleum, energy, paper, grain and research/education sectors.
To learn more about moisture analyzers, follow /, or leave a message or call our customer service team.
Recently Viewed Products
Copyright © Jiangsu Victor Instrument Meter Co., Ltd. All rights reserved | Sitemap | Technical Support: VicoMeter Operations Center