Anhui Haihua Technology Co., Ltd.
Integrating multiple advanced technologies, the VicoMeter WKT-A6 fully automatic trace moisture analyzer works on the internationally recognized classic Karl Fischer coulometric method. Its measuring range goes down to 0.0001% (1 ppm), and it measures moisture in liquids, solids, and gases within 60 seconds. It is widely used in chemicals, petroleum, power, research, railways, and other industries where moisture content is one of the key quality indicators. Multiple built-in moisture formulas meet different users' needs; the efficient mini printer and large built-in memory store 500 test records, with convenient retrieval so users can easily view and print test reports.
Coulometric (coulometry) principle, ideal for trace-level testing; 7-inch color touchscreen with smart operation, automatic result calculation, and freely configurable sample IDs; complies with multiple national standards and suits virtually every industry.
8 moisture calculation formulas, 6 determination methods, and 8 end-point detection algorithms
Automatic drift compensation, automatic subtraction of ambient and background moisture, and self-diagnostic fault alerts
One-touch start for reliable results
Outstanding automation — connects seamlessly to the Karl Fischer oven for difficult samples
RS232 interface for connection to a PC, printer, PLC, or LIMS
The customer had owned the instruments for some time and, having used similar equipment before, did not request on-site installation. The VicoMeter engineer's visit this time was prompted by small problems with both instruments: on the older one the stirrer would not spin, and the newer one could never stop equilibrating.
On site, the engineer first checked both instruments and found the electrodes and titration cells were quite dirty. During cleaning, a black deposit remained on the platinum mesh of the generating electrode of the instrument that would not equilibrate; methanol would not wash it off, so the engineer borrowed the customer's ultrasonic cleaner to clean it. On the other instrument, the stir bar turned out to be larger than normal; a smaller one was fitted and stirring returned to normal.
The engineer then dealt with the second instrument: since the two were identical, the good electrode was swapped onto the instrument that would not equilibrate — after the swap it reached electrolysis equilibrium in under ten minutes, confirming a faulty electrode. The instruments come with stir bars and other accessories matched at the factory, so avoid replacing them arbitrarily. The platinum electrode is a critical component — clean it every time the reagent is changed to extend the instrument's service life.
Anhui Haihua Technology Co., Ltd. is a large integrated technology-industry-trade chemical enterprise specializing in intermediates for pharmaceuticals, pesticides, food additives, and household chemicals, with its own import/export rights. Bordering the Beijing-Shanghai railway to the south and the Nanjing-Luohe expressway to the west, the plant enjoys an ideal location with convenient transport, covering 270,000 m² with 65,000 m² of building floor space. With advanced production facilities and excellent testing equipment, it is certified to ISO 14001:2004 (environmental) and ISO 9001:2008 (quality), and has developed and put into production more than ten fine chemical products, mainly m-cresol, p-cresol, m/p-cresol, o-cresol, 2,6-di-tert-butyl-p-cresol, xylyl ether (high-temperature heat transfer oil), thymol, m-nitrochlorobenzene, m-chloroaniline, and 6-tert-butyl-m-cresol.
Related Products
Related Knowledge
Related Cases
Customized solutions for your real business needs
ContactRecently Viewed Products
Copyright © Jiangsu Victor Instrument Meter Co., Ltd. All rights reserved | Sitemap | Technical Support: VicoMeter Operations Center