Lingyun Automotive Hose Co., Ltd. purchased the VicoMeter WKT-A9 precision Karl Fischer moisture analyzer. The VicoMeter WKT-A9 precision Karl Fischer trace moisture analyzer uses the coulometric Karl Fischer principle to quickly determine trace moisture in samples, measuring water content as low as 0.001 ppm. Guided navigation prompts and a built-in electronic manual make operation simple. The WKT-A9 complies with multiple national and international moisture determination standards, meets the moisture measurement needs of various industries, and is widely used in chemicals, petroleum, pharmaceuticals, electric power, food, research, testing and other industries—a Karl Fischer moisture analyzer for mid-to-high-end applications.

WKT-A9 Karl Fischer Trace Moisture Analyzer Coulometric

WKT-A9 Karl Fischer Trace Moisture Analyzer Coulometric

WKT-A9 Karl Fischer Trace Moisture Analyzer Coulometric

Coulometric (coulometry) principle, ideal for trace-level testing; 5-inch color touchscreen with smart operation, automatic result calculation, and freely configurable sample IDs; complies with multiple national standards and suits virtually every industry.

  • 8 moisture calculation formulas, 6 determination methods, and 8 end-point detection algorithms

  • Automatic drift compensation, automatic subtraction of ambient and background moisture, and self-diagnostic fault alerts

  • One-touch start for reliable results

  • Outstanding automation — connects seamlessly to the Karl Fischer oven for difficult samples

  • RS232 interface for connection to a PC, printer, PLC, or LIMS


VicoMeter at the Customer Site

Lingyun Automotive Hose Co., Ltd. purchased the VicoMeter WKT-A9 Precision Karl Fischer moisture analyzer mainly to test the moisture content of their lubricating oil. After arriving at the customer's site, the VicoMeter technical engineer first unpacked and checked the instrument and its accessories with the customer, and then began the installation.

Installation was completed quickly. Afterwards, the VicoMeter technical engineer verified the instrument with 0.1 uL, 0.2 uL, 0.5 uL and 2 uL of deionized water; all results were within the permitted range. The customer then brought samples for the engineer to test: one previously taken and stored in a sealed glass bottle, and one freshly taken in a plastic bottle on site.

After receiving the samples, the VicoMeter technical engineer shook them thoroughly to homogenize, rinsed the syringe three times, then drew samples for testing. The sample from the sealed glass bottle was tested three times, with moisture around 150 ppm each time. The freshly collected sample in the plastic bottle also gave consistent results, but the values were around 1200 ppm—clearly much higher. The engineer then went to the customer's site, took a new sample personally in a new sealed glass bottle and ran two more tests; this time the results were around 130 ppm, clearly much better than the customer's earlier samples.

The VicoMeter technical engineer analyzed the results; the main reasons were:

1. The customer did not use dedicated sampling bottles

2. The customer did not rinse the bottle with the sample before sampling

3. The bottle was not filled completely, and the sample absorbed moisture from the air.

Together, these reasons explain why the customer's samples showed significantly higher moisture than the engineer's; sampling practice also has a considerable influence on the results.

Customer Profile

WKT-A9 Karl Fischer Trace Moisture Analyzer Coulometric

Lingyun Automotive Hoses

(Click image to enlarge)

Lingyun Automotive Hoses

Kuadan-Lingyun Automotive Hose Co., Ltd. is a wholly owned subsidiary of Hebei Lingyun Industrial Group under the China North Industries Group. Founded in 1995, it specializes in automotive air conditioning hoses, fuel hoses, power steering hoses, shaped hoses, oil cooler hoses, industrial hoses and hose assemblies, and develops other related rubber products, holding a leading position in the industry. The company imported a complete set of internationally advanced European hose and assembly production and testing equipment, and obtained national laboratory accreditation in 2009, with equipment precision and testing capabilities at an internationally advanced level. It has an overseas technical center, relying on Danish Kuadan technology R&D personnel to keep pace with advanced European rubber hose technology, ensuring the 'high quality, high starting point' principle and product technology in step with the world. The company is certified to ISO/TS 16949:2002, ISO 14001:2004, OHSAS 18001:1999, ISO 9001 and CCAP, holds national laboratory accreditation and Hebei high-tech enterprise recognition, and has been named an outstanding supplier by many customers for consecutive years. The company is in a period of rapid growth, with output value expected to exceed 700 million yuan under the 12th Five-Year Plan.

Related Products

Related Knowledge

Karl Fischer Moisture Analyzers Karl Fischer Moisture Analyzers

A Karl Fischer moisture analyzer is an instrument designed on the Karl Fischer method to measure moisture content in materials. It is also known as a Karl Fischer moisture meter, fully automatic Karl Fischer moisture titrator, intelligent Karl Fischer trace moisture analyzer, or Fischer-method moisture meter, and comes in two types: coulometric and volumetric (titrimetric).

Which Samples Can a Karl Fischer Moisture Analyzer Test? Which Samples Can a Karl Fischer Moisture Analyzer Test?

As a classic moisture determination method, Karl Fischer titration is suitable for measuring water in a wide range of organic and inorganic substances. Because compounds vary in their properties, they fall into two categories: those that can be analyzed directly and those that cannot.

What Is the Difference Between Coulometric and Volumetric Karl Fischer Titration? What Is the Difference Between Coulometric and Volumetric Karl Fischer Titration?

The biggest difference between the volumetric and coulometric methods is that I2differ in their source: in the volumetric method, I2comes from the titrant reagent, while in the coulometric method, I2is produced by electrolyzing an electrolyte containing I-is generated by electrolysis of the electrolyte.

Related Cases

Zhejiang Xinjin Air Conditioning Equipment
Nantong Yamaguchi Precision Electromechanical
Guangzhou Muqiao Auto Parts Co., Ltd.
Lingyun Automotive Hoses

Customized solutions for your real business needs

Contact

Recently Viewed Products

Contact
+8652389989770
+8613852641385
sales@vicometer.com