Nantong Shankou Precision Electromechanical purchased the VicoMeter WKT-A9 Precision Karl Fischer moisture analyzer. The VicoMeter WKT-A9 Precision Karl Fischer trace moisture analyzer uses the coulometric Karl Fischer (coulometry) principle to rapidly measure trace moisture in samples, detecting water down to 0.001 ppm and determining trace moisture in solids, liquids, and powders within one minute. The built-in printer and storage for 500 data sets allow real-time query and printout. Innovative guided prompts and a built-in e-manual make operation simple. The WKT-A9 complies with multiple national and international moisture standards and provides eight built-in moisture calculation formulas and six determination methods to meet the moisture testing needs of every industry. Widely used in chemicals, petroleum, pharmaceuticals, power, food, research, and testing, it is a Karl Fischer moisture analyzer built for mid-to-high-end applications.
Coulometric (coulometry) principle, ideal for trace-level testing; 5-inch color touchscreen with smart operation, automatic result calculation, and freely configurable sample IDs; complies with multiple national standards and suits virtually every industry.
8 moisture calculation formulas, 6 determination methods, and 8 end-point detection algorithms
Automatic drift compensation, automatic subtraction of ambient and background moisture, and self-diagnostic fault alerts
One-touch start for reliable results
Outstanding automation — connects seamlessly to the Karl Fischer oven for difficult samples
RS232 interface for connection to a PC, printer, PLC, or LIMS
When the VicoMeter service engineer arrived on site, the ambient humidity was very high — as much as 78%. Following the engineer's advice, the customer closed the doors and windows and switched on the air-conditioning dehumidifier; after about an hour the laboratory humidity dropped to 55%, meeting the instrument's installation requirements.
After installation, the engineer began testing the customer's samples — an additive used in cleaning agents with low moisture content, no aldehydes or ketones, and neutral pH. A standard pyridine-free coulometric Karl Fischer reagent was therefore selected. Because the sample's water content was low, a 2 mL syringe was used, since a larger sample volume is needed to guarantee accurate data. A 1.9856 g sample gave 102 ppm and a 2.0152 g sample gave 108 ppm — stable, repeatable results that fully satisfied the customer.
When dealing with samples containing very little water, appropriately increasing the sample size makes the results more stable and the data more reliable.
Founded on April 20, 2005 and registered in the industrial zone of Pingchao Town, Tongzhou District, Nantong City (Yuntaishan Village), with Xu Ti as legal representative, Nantong Shankou Precision Electromechanical Co., Ltd. manufactures and sells electromechanical equipment, bearings, and computer components, along with product R&D and self-run import/export business (subject to approval where required). The company holds equity in two other companies.
Related Products
Related Knowledge
What Calibration Standards and Verification Methods Are Available for Karl Fischer Moisture Determination?
Related Cases
Customized solutions for your real business needs
ContactRecently Viewed Products
Copyright © Jiangsu Victor Instrument Meter Co., Ltd. All rights reserved | Sitemap | Technical Support: VicoMeter Operations Center